background image

 

Start 

 

Place the scale, and using the bubble on the backside for level It. 
Press On / Off to turn on the scale. Following a countdown, marked also by the beep, you will see the 
software version, the capacity of the scale, and then you enter the measurement mode. 

Wait at least 30 minutes to warm up before the first use. 
 

Date and Time settings

 

 

Switch on the balance. Keep pressed 

 for 3 seconds for date and time settings menu: “

rtc-set”

First value requested is the year: the display shows “

Year **”: 

Use the following buttons to modify the blinking value: 

 

  

 set the value, 

 

  

 scroll to the next value   

 

  

 confirm. 

 Successively are required: 

1.

 

Month   

Month

 

2.

 

Day 

 

Day

 

3.

 

Hour 

 

Hour

 

4.

 

Minute   

Min

 

User Calibration 

 

Each balance is calibrated in the laboratory by technicians using certified sample masses. 
Provided with all models there is a sample mass useful to test the functionality of the instrument

 

Calibration procedure

:

 

 

o

 

Turn on the balance and wait at least 30 min. of warmup 

o

 

Keep pressed 

 button for 3 seconds, 

the display shows first “

Cal 0

”, “

-------

“, and then it shows the sample mass [g] 

requested for the calibration. * 
 

o

 

Place the sample mass on the plate. 
 

o

 

Display first shows “-------“, than the weight measurement. 
 

Calibration is complete. 
 

 

* With the 

 button is possible to change this value, choosing one of the allowed for that model (Tab 1) 

Summary of Contents for BL 303

Page 1: ...User Manual BL 303 BL 602 BL 2002 BL 3001 BL 6001 BL 30K1 Models...

Page 2: ...START 7 DATE AND TIME SETTINGS 7 USER CALIBRATION 7 CALIBRATION PROCEDURE 7 UNIT SETTING 8 UNIT SELECTION 8 DEFAULT UNIT SETTINGS 8 ZERO 8 TARE 8 MEASURE 8 COUNTER FUNCTION 9 RS 232 SETTINGS 9 ITEM SE...

Page 3: ...3 Glass protection BL 303...

Page 4: ...4 BL 602 1 2 3 4 5 6 7 8...

Page 5: ...5 BL 2002 BL 3001 BL 6001...

Page 6: ...Working Humidity 40 85 Installation and removal of the scale must be done with the machine switched off Do not use the scale upside down After switch On wait at least 30 minutes warming up before the...

Page 7: ...e value scroll to the next value confirm Successively are required 1 Month Month 2 Day Day 3 Hour Hour 4 Minute Min User Calibration Each balance is calibrated in the laboratory by technicians using c...

Page 8: ...ith no weight on the scales press the button to take the 0 00 reading on the display The ZERO function is only effective if the value shown is less than 4 of total capacity If the value is greater tha...

Page 9: ...he display shows first rS Set 8888888 According with the following table set the correct values a 0 9600 1 19200 b 0 print in continuos 1 print when stable 2 press to print c 0 1 print the date d 0 1...

Page 10: ...trument in correspondence of the load cell Datasheets Models BL 303 BL 602 BL 2002 BL 3001 BL 6001 BL 30K1 Cod 27000513 27000323 27000313 27000113 27000123 27000133 Capacity 300 g 600 g 2000 g 3000 g...

Page 11: ...of in public landfills According to the UE Directive 2002 96 EC the European users of electrical and electronic equipment can return it to the dealer or manufacturer upon purchase of a new one The il...

Reviews: