TriOxmatic
®
690
Maintenance, cleaning, disposal, and replacement
17
ba25314e05
09/2020
The ELY/A electrolyte solution irritates eyes, skin and mucus
membranes. If it comes into contact with the eyes, rinse
thoroughly with water and consult a doctor! During working
activities, always wear suitable protective gloves and protective
goggles/face shield! Follow the safety datasheet.
5
Unscrew the membrane cap (for disposal of
membrane cap and electrolyte solution, see section 5.7).
6
Rinse the sensor head with tap water.
7
Carefully rub the silver counter electrode with a paper towel
and rinse it with deionized water.