Firmware update
Turb 430 IR/T
58
ba75507d09
08/2021
11
Firmware update
General information
With the "Firmware Update Turb430" program you can update the firm-
ware of the Turb 430 IR/T to the latest version with the aid of a PC.
A free serial interface (COM port) on your PC and an interface cable is
required for this (see chapter 8 A
).
Program installation
Install the firmware update program on your PC with the "Turb430_Vx-
_yy_English.exe" installation program.
Program start
Start the "Firmware Update Turb430" program from the WTW directory
in the Windows start menu. The program automatically selects the first
free serial interface (COM port). The selected interface is displayed on
the left side of the status line on the screen bottom.
Via the language menu you can change the adjusted language.
Firmware update
Proceed as follows:
After switching the meter off and on you can check whether the meter
has taken over the new software version on the start display.
Prior to starting the update please make sure that the bat-
teries are fully loaded, or operate the Turb 430 IR/T with
the LabStation or with the power pack. Otherwise there is
the risk of the Turb 430 IR/T crashing during the update.
1 With the aid of an interface cable, connect the Turb 430 IR/T to
the serial interface (COM port) of the PC named in the status
line.
2 Make sure the Turb 430 IR/T is switched on.
3 To start the updating process click the OK button.
4 Then follow the instructions of the program.
During the programming process, a corresponding message
and a progress bar (in %) appear.
The programming process takes approx. four minutes.
A terminatory message is displayed after a successful pro-
gramming process. The firmware update is now completed.
5 Disconnect the meter from the PC.
The meter is ready for operation.