10

Figure 7

  ProQuatro Keypad

2.1

Keypad Layout

1

5

6

7

8

9

10 11

2

3

4

2.  Operation

1

System:

 Opens System Menu from any screen. Use to adjust system settings.

2

Probe:

 Opens Sensor Menu from any screen. Use to enable sensors and display units.

3

Calibrate:

 Opens Calibrate Menu from any screen. Use to calibrate all parameters except temperature.

4

File:

 Opens File Menu from any screen. Use to view data and calibration records, set up Data IDs, and delete data.

5

Exit/Escape key:

 Exits back to Run Screen. When in alpha/numeric entry screen, escapes to previous menu.

6

Backlight:

 Press to turn the instrument backlight on and off and to adjust the display contrast when pressed with the left or 

right arrow key.

7

Left arrow key:

 Use to navigate left in alpha/numeric entry screens. Press to return to previous menu in all screens except 

alpha/numeric entry. Can be pressed simultaneously with Backlight key to decrease display contrast.

8

Right arrow key:

 Use to navigate right in alpha/numeric entry screens. Can be pressed simultaneously with Backlight key to 

increase display contrast.

9

Power: 

Press to turn the instrument on. Press and hold for 5 seconds to turn off.

10

Up/Down arrow keys:

 Use to navigate through menus and to navigate down in alpha/numeric entry screens.

11

Enter key:

 Press to confirm selections, including alpha/numeric key selections. When on the Run screen, pressing the Enter 

key stores data.

Summary of Contents for 606950

Page 1: ...ProQuatro User Manual PROFESSIONAL SERIES HANDHELD METER USER MANUAL 606962 01...

Page 2: ...ation direct incidental and consequential damages and disclaims such damages to the full extent permitted under applicable law The user is solely responsible to identify critical application risks and...

Page 3: ...Chloride Calibration Barometer Calibration Restore Default Calibration Values Recalibration Prompt 2 7 Files Menu Data Memory Viewing Saved Data View Calibration Records View Data ID List Delete Data...

Page 4: ...he instrument over to view the battery cover on the back 2 Unscrew the four captive battery cover screws 3 Remove the battery cover and instal the new batteries ensuring correct polarity alignment Fig...

Page 5: ...0 ISE Cond Temp Includes built in conductivity and temp sensors There is one ISE port in which an ISE can be installed ISE sensors sold separately 6051020 1 4 10 20 or 30 DO ISE Temp Includes a built...

Page 6: ...DO installation then port 2 and lastly C T Cable Port 1 Options Port 2 Options 1010 dual port cable pH pH ORP ORP Ammonium Ammonium Chloride Chloride Nitrate Nitrate None use a port plug Cable Port 1...

Page 7: ...oxygen sensor and connecting the cable to the ProQuatro the ProQuatro must be configured for the sensor being installed See the Dissolved Oxygen Sensor Setup section of this manual for instrument con...

Page 8: ...the nut down completely without feeling any resistance Damage to your cable sensor may occur if you force the parts together 6 Once completely installed the nut will seat flat against the bulkhead At...

Page 9: ...emp single junction combination electrode with 1 meter cable 605179 pH ORP Temp single junction combination electrode with 4 meter cable The military spec MS cable connectors are keyed for positive ma...

Page 10: ...acklight on and off and to adjust the display contrast when pressed with the left or right arrow key 7 Left arrow key Use to navigate left in alpha numeric entry screens Press to return to previous me...

Page 11: ...n indicator This is only shown when sending data to a USB flash drive 4 Scroll bar 5 Message area 6 Displayed measurements 7 Sampling mode indicator Log One Sample is displayed when One Sample Logging...

Page 12: ...Time from the System menu Press enter to select Date Time Options Date Format Highlight and press enter to open a sub menu for selecting the preferred date format YY MM DD MM DD YY DD MM YY or YY DD...

Page 13: ...de Press ENTER to begin logging Logging is stopped by pressing ENTER again The handheld will beep when logging is started and stopped One sample logging Clear the Continuous Mode check box The Run scr...

Page 14: ...h drive included with the instrument is FAT32 formatted DO NOT place the firmware file within any folder on the flash drive 3 Highlight Update Firmware and press Enter 4 Connect the USB flash drive to...

Page 15: ...sed Highlight Enabled and press Enter to activate checked box or deactivate unchecked box the DO channel Sensor Type sets the type of oxygen sensor being used either Polarographic black body or Galvan...

Page 16: ...vity channel Temp Ref Temperature Reference is the reference temperature used for calculating temperature compensated Specific Conductance This will be the temperature all Specific Conductance values...

Page 17: ...tely from specific conductance unless the make up of the chemical species in the water remains constant A To change the multiplier highlight TDS Constant and press enter Use the numeric entry screen t...

Page 18: ...an be displayed in uS cm or mS cm Conductivity is the measure of a solution s ability to conduct an electrical current Unlike specific conductance conductivity is a direct reading without any temperat...

Page 19: ...ISE2 Display 100 Cl mg L Cl mV pH Display Press the Probe key highlight Display and press enter Highlight ISE pH and press enter You will not be able to Display the sensor unless it is Enabled in the...

Page 20: ...ay in real time For example if DO and pH have Auto Stable and Hold All Readings enabled then all readings currently on the display e g specific conductance temperature ORP in addition to DO and pH wil...

Page 21: ...is is the value that should be observed to ensure the measurement is stable before choosing Accept Calibration Post Cal Value This is the same as the calibration value This will be the measurement val...

Page 22: ...led A ring stand should be used to support this container during calibration YSI recommends calibrating conductivity first as the salinity reading from the conductivity sensor is a variable used to de...

Page 23: ...libration If using the cup and you no longer have the sponge place a small amount of clean water 1 8 inch in the plastic storage cup instead Make sure there are no water droplets on the DO membrane or...

Page 24: ...the Calibrate Menu highlight DO and press enter Select Zero The Calibration value is automatically adjusted to 0 so there is no need to modify this value Observe the Actual Readings for stability no...

Page 25: ...olutions Additional Calibration Points Up to a 6 point calibration can be completed if the buffer set selected under ISE Sensor Setup is set to Off This allows for any combination of buffers to be use...

Page 26: ...selected or the pH sensor has been calibrated to 6 calibration points ORP Calibration Obtain a premixed standard solution that is approved for use with Ag AgCl ORP sensors or prepare a standard with...

Page 27: ...ht the appropriate ISE and press enter Select Calibration value and enter the value that corresponds to the first calibration standard Observe the Actual Readings for stability no significant change f...

Page 28: ...meter is factory calibrated and should rarely need to be recalibrated The barometer is used for DO calibration and DO measurements Verify that the barometer is accurately reading true barometric press...

Page 29: ...r calibration has not occurred within the user defined number of days the instrument will display a short message reminding the user to calibrate the specified channels Figure 46 Re Cal Prompts 03 03...

Page 30: ...eys to scroll through the data Data ID View data from one ID or all IDs Begin End View data within specific date and time ranges Figure 47 Files Menu 03 03 20 10 34 42 AM Files 100 Data Memory free 10...

Page 31: ...ensor Setup menu TDS Constant User selected in Sensor Setup menu Temperature Cal Cell Constant Calibrate Status DO Sensor Calibration Record Content Date and Time Method mg L Cal Value Sensor Value Se...

Page 32: ...entries can be created by choosing Add new Data IDs already created are shown in alphabetical order on the Data ID list Choose an entry from the Data ID list to Select Edit or Delete When selected da...

Page 33: ...connector After connecting a flash drive to the instrument via the adapter cable and the USB symbol has appeared under the battery indicator highlight Upload Files and press Enter Once Upload Success...

Page 34: ...table parameters if applicable 6 On the main run screen press ENTER to begin logging single or interval continuous See System Menu g Logging for more information a An option to change the Data ID if e...

Page 35: ...the o ring should be replaced with an identical o ring At the time of o ring replacement the entire o ring assembly should be cleaned To remove the o rings Use a small flat bladed screwdriver or simi...

Page 36: ...o the o ring Place the o ring into its groove making sure that it does not twist or roll Use your grease coated finger to once again lightly go over the mating surface of the o ring Sensor Ports It is...

Page 37: ...ay appear tarnished and operate just fine YSI recommends using the 400 grit wet dry sanding disks to resurface the electrodes if the sensor has difficulty stabilizing or calibrating after a membrane c...

Page 38: ...wet paper towel to remove the residual layer from the anode You can smell the tip of the sensor to help ensure all the ammonia has been rinsed off Trapping residual ammonia under the new membrane cap...

Page 39: ...s based on membrane type typically form more rapidly with the 5912 1 mil Teflon less rapid with 5913 1 25 mil PE and least rapid with 5914 2 mil PE At the time the membrane cap is changed YSI recommen...

Page 40: ...sits or contaminants appear on the glass and or platinum surfaces or when the sensor s response slows Removing the sensor from the cable may make cleaning easier To chemically clean the sensor 1 Soak...

Page 41: ...an be regenerated by washing with alcohol and or gently polishing with fine emery paper in a circular motion to remove any deposits or discoloration then thoroughly washing with deionized water to rem...

Page 42: ...onductivity electrodes are not corrosive for example chlorine bleach However it is recommended that the sensor be cleaned with the provided brush prior to and after long term storage The storage tempe...

Page 43: ...e sensor should remain submerged in the solution during the storage period therefore make certain that the vessel is sealed to prevent evaporation and periodically check the vessel to ensure the senso...

Page 44: ...Incorporated 1725 Brannum Lane Yellow Springs OH 45387 USA Web Visit YSI com to order replacement parts accessories and calibration stolutions When placing an order please have the following available...

Page 45: ...d Temp Includes built in conductivity and temp sensors There is one DO port in which a DO sensor can be installed DO sensors sold separately 6051030 1 4 10 20 or 30 ISE Cond Temp Includes built in con...

Page 46: ...MS connector Lab Cable Assemblies features built in cables and sensors Other Accessories YSI Item Description 603075 Carrying Case Soft Sided 603074 Carrying Case Hard Sided 603162 Carrying Case Soft...

Page 47: ...ctivity Calibrator 50 000 umhos cm 8 ea pint 061320 Zobell Solution ORP Calibrator 125 mL 061321 Zobell Solution ORP Calibrator 250 mL 061322 Zobell Solution ORP Calibrator 500 mL 003821 pH 4 Buffer b...

Page 48: ...oughout the United States and Internationally For the nearest service center information please visit ysi com and click Support or contact YSI Technical Support directly at 800 897 4151 1 937 767 7241...

Page 49: ...05177 605178 605179 605202 605203 605323 605324 605780 Cables 60510 xx 60520 xx 60530 xx 6051010 xx 6051020 xx 6051030 xx 6052030 xx Accessories 626444 Conforms to the following Directives EMC 2014 30...

Page 50: ...ne Yellow Springs OH 45387 USA Product Name ProQuatro Water Quality Instrument Model Number ProQuatro 606950 Sensors 605107 605108 605109 605177 605178 605179 605202 605203 605323 605324 605780 Cables...

Page 51: ...ance of the original warranty period or at least 90 days from date of repair or replacement LIMITATION OF WARRANTY This Warranty does not apply to any YSI product damage or failure caused by 1 Failure...

Page 52: ...93 684 0 91 19 911 93 89 26 63 676 4 90 18 901 79 88 26 33 668 8 89 17 891 66 87 26 03 661 2 88 15 881 53 86 25 73 653 6 87 14 871 40 85 25 43 646 0 86 13 861 26 84 25 13 638 4 85 11 851 13 83 24 83 6...

Page 53: ...9 58 9 02 8 0 11 84 11 17 10 53 9 93 9 36 8 83 9 0 11 56 10 91 10 29 9 71 9 16 8 64 10 0 11 29 10 66 10 06 9 49 8 96 8 45 11 0 11 03 10 42 9 84 9 29 8 77 8 28 12 0 10 78 10 18 9 62 9 09 8 59 8 11 13...

Page 54: ...55 35 0 6 95 6 62 6 31 6 02 5 73 5 46 36 0 6 84 6 52 6 22 5 93 5 65 5 38 37 0 6 73 6 42 6 12 5 84 5 56 5 31 38 0 6 62 6 32 6 03 5 75 5 48 5 23 39 0 6 52 6 22 5 98 5 66 5 40 5 15 40 0 6 41 6 12 5 84 5...