Commissioning
Alyza IQ PO4
44
ba76201d03
05/2020
3.3.10 Mounting the cover plate for the control unit ACM
The cover plate for the control unit ACM covers the control unit ACM and the ca-
bles connected to it.
figure 3-15Cable of the control unit ACM
1
3-pole connection (mains)
2
4-pole connection (data)
3
USB connection
4
RJ45 connection
1
Remove the transport protection of the measuring unit (see
section 3.3.8 Removing the transport protection of the measuring unit,
42).
4 cable ends are on the measuring unit.
2
Connect the 4 cables lying on the measuring unit to the control unit (see
section 3.3.9 Connecting the cables to the control unit ACM,
3
Unscrew the 2 knurled-head screws from the top hat rail on the housing
top,
4
Screw the cover plate to the top hat rail with the 2 knurled-head screws.
3 4
2
1
Summary of Contents for Alyza IQ PO4-XYZ