Kalibrierhistorie
The
Kalibrierhistorie
list shows the last calibration results (see section 4.4.3 Cal-
figure 4-10Kalibrierhistorie Alyza IQ
The calibration history provides the following information:
Date
and
time
of the calibration
Srel
(relative slope, is determined with 2-point-calibration)
Value without asterisk (*):
the slope was determined (2-point calibration)
* value with asterisk (*) :
the slope was not determined (1-point calibration);
the last active slope was taken over unchanged
OS
(offset, is determined with 1- and 2-point calibration)
Type
M: Manual calibration
A: Automatic calibration
F: Factory setting
STD
(concentration in mg/l of the calibration standards used)
MR
(measuring range 1 [low] or 2 [high])
OK
(evaluation of the calibration result):
+ successful calibration.
The new calibration data are taken over for measurement.
- calibration unsuccessful.
Sensor blocked for measurement.
? : The last valid calibration or
factory setting was taken over
Current
calibration data
Chronological list of
the last calibration
procedures