Measurement / Operation
Alyza IQ PO4
98
ba76201d03
05/2020
4.7
Software update for the Alyza IQ
With a Software-Update you can alway update your Alyza IQ to the latest status
of the instrument software.
For the update packet with the current instrument software for the Alyza IQ as
well as a comprehensive manual on how to carry out the update, go to
www.YSI.com.
The Alyza IQ is an IQ S
ENSOR
N
ET
component with separate USB interface.
5
Select the function,
Service-Dateien auf USB-Speicher übertragen
.
Follow the instructions on the display.
6
Remove the USB memory device from the interface "USB0".
7
If necessary, re-insert the previously removed USB plug into the unla-
beled USB interface.
8
Reinsert the cover and fix it with the 2 fixing screws.
9
Connect the USB memory device to the USB interface of a PC and dis-
play the file system of the USB memory device.
On the USB memory device there is the folder "LogData" with the oper-
ating data of the Alyza IQ.
The folder includes several files with the operating data of the Alyza IQ.
The assignment of the data to an Alyza IQ is secured via the series
number of the Alyza IQ in the file name.
10
If requested, send some individual files or the entire folder "LogData" to
the service department.
The Software-Update for the Alyza IQ
is included in the update
packet for the IQ S
ENSOR
N
ET
.
The software versions of all components can be viewed in the dialog
box
Liste aller Teilnehmer
(see IQ S
ENSOR
N
ET
system operating
manual).
1
Download the software update IQ S
ENSOR
N
ET
"Update Pack (L1)" and
store the directories to a USB memory device.
2
Unscrew the 2 fixing screws of the cover lid of the control unit ACM and
remove the cover lid of the ACM.
Summary of Contents for Alyza IQ PO4-XYZ