Alyza IQ PO4
149
ba76201d03
05/2020
5.12 Transport, storage
5.12.1 General notes
Please observe the allowed environmental conditions for transport and storage
of the Alyza IQ (see section 7.3 General data,
Please not that no liquids or ChemBags may be in the Alyza IQ for storage and
transport.
NOTE
Inappropriate transport and inappropriate storage of the Alyza IQ (especially of
the measuring unit) can cause damage due spilled liquids.
Safe transport and safe storage are only possible if the Alyza IQ is free from liq-
uids (tubes, photometer unit, overflow vessels empty; ChemBags removed).
5.12.2 Preparing the Alyza IQ for transport or storage
Preparing the
measuring unit for
transport or
storage
18
Now carry out the intended activities, such as:
Dismount the measuring unit and carry out maintenance activities
(see section 5.10 Complex maintenance- and cleaning activities in
the measuring unit,
Prepare and carry out storage or transport
(see section 5.12 Transport, storage,
Even minor damage to tubes or ChemBags due to mechanical
stress (e.g. pressure, shock etc.) or thermal impact (e.g. frozen liq-
uids) can cause consequential damage due to spilled liquids.
1
Decommission the Alyza IQ (see section 5.11 Decommissioning,
2
Open the front cover of the measuring unit.
3
Remove all ChemBags from the supporting rod.
4
Remove the cover of the control unit ACM.
5
Unplug the cables from the switch box.
6
Close the measuring unit.
7
Fix the measuring unit in its position inside the housing of the Alyza IQ
with the 3 original transportation safety devices.
Summary of Contents for Alyza IQ PO4-XYZ
Page 171: ......