Alyza IQ PO4
Measurement / Operation
79
ba76201d03
05/2020
– Activate or terminate the maintenance condition
– Start calibration
– Start the service functions
4.2
Measurement operation
4.2.1 Determination of measured values
The Alyza IQ determines the measured values with a chemical analyzing proce-
dure. The measured values are determined at intervals specified in the settings
(see section 4.3 Settings for the Alyza IQ,
81).
If the intervals overlap with the specified start times and intervals of other func-
tions (
Autom.Reinigung
or
Autom.Kalibrierung
), the functions are carried out ac-
cording to their priority (see section 4.3.2 Priority,
85).
4.2.2 Starting the measuring operation
The Alyza IQ starts measuring only when the
Alyza IQ STARTEN
function is
started.
When the Alyza IQ is started (function
Alyza IQ STARTEN
),
it checks automat-
ically whether the requirements for measurement are met, e.g.:
Plausible pressure and volume conditions in the tubes
Temperature control of the various areas of the Alyza IQ
The regulating of the temperature may take up to 30 minutes.
More details: See section 3.4.4 Preparing the Alyza IQ for measuring,
4.2.3 Measuring
The measured values are displayed in the measured value display on the termi-