BA80009def04
33
English
Français
- Even if no condensation could be
found in the data logger housing: pull
the battery plug from the socket and
wait 30 seconds.
- Then reconnect the battery and
check that the LED to the right of the
battery compartment is flashing. If
the LED does not flash, the battery
may be flat and needs to be replaced
(see “Replacing the Battery”).
- As a test, insert a fresh battery in the
socket
on
the
circuit
board.
If the LED still does not flash (despite
the fresh battery), the problem lies
with the data logger.
- Before you close the data logger
housing again, replace the silicone
O-ring.
A used silicone O-ring must always
be
disposed
of.
Under
no
circumstances should it be reused!
This is because it is deformed when
the two parts of the housing are
screwed together. If it is used again it
will not make a proper seal!
-
Même si vous n’avez pas trouvé
d’humidité dans le boîtier de
l’enregistreur : débranchez la prise
de la batterie et attendez 30
secondes.
- Rebranchez ensuite la batterie et
vérifiez si la LED située à la droite du
compartiment
de
la
batterie
recommence à clignoter. Si la LED
ne clignote
pas, cela signifie
probablement que la batterie est vide
et
doit
être
remplacée
(cf.
paragraphe « Remplacement de la
batterie »).
- Pour tester, branchez une nouvelle
batterie dans la prise femelle de
l’appareil.
Si la LED ne clignote toujours pas en
dépit du remplacement de la batterie,
il est probable que le problème
vienne de l’enregistreur.
- Avant de refermer le boîtier de
l’enregistreur,
procédez
systématiquement au remplacement
du joint torique d’étanchéité en
silicone.
Tout joint torique d’étanchéité en
silicone utilisé une fois doit être jeté:
il ne doit être en aucun cas réutilisé !
Lors du vissage des deux parties du
boîtier, le joint est en effet déformé et
ne peut donc pas assurer une
véritable
étanchéité
lors
d’une
seconde utilisation !