• Check from time to time that the salt
solution is still saturated and that there
is still liquid present. To do this remove
the bottle from the foam block and tip it
slightly (the presence of liquid must still
be apparent). If this is not the case add
demineralised water (distilled water) in
small quantities. Wait 48 hours before
renewed use of the equipment.
• The bottles must always be kept
closed with the original caps insofar as
they are not in use.
• The upper parts of the bottles must not
come into contact with water or salt.
• When a sensor with a moisture probe
is introduced into a bottle with a salt
solution, it must be ensured:
- that the sensor is not damaged.
- that the sensor does not come into
contact with the salt solution.
• Stabilisation is reached after approx. 2
hours following opening of the bottle and
introduction of the probe.
• Vérifiez de temps en temps si la solu-
tion
saline n’est pas encore saturée et
s’il y a encore du liquide. Pour ce faire,
retirer le flacon de son bloc de mousse
et inclinez-le légèrement (vous devez
constater la présence de liquide). Si ce
n’est pas le cas, ajouter de petites
quantités d’eau déminéralisée (eau
distillée). Atte,dre 48 heures avant toute
nouvelle utilisation du dispositif.
• Dans la mesure où ils ne sont pas
utilisés, les flacons devraient être
maintenus bouchés avec leur capuchon
d’origine.
• La partie supérieure des flacons ne
doit
pas entrer en contact avec de l’eau
ou du sel.
• Lorsque vous introduisez le capteur
avec le palpeur d’humidité dans le
flacon, vous devez veiller
- à ne pas endommager le capteur.
-
à ce que le capteur n’entre pas en
contact avec la solution saline.
• Une fois le flacon ouvert et le capteur
introduit, la stabilisation intervient au
bout de 2 heures environ.