pH value
Multi 3620 IDS
40
ba77169d06
10/2021
Calibration record
(USB output)
5.3
Continuous measurement control (CMC function)
The Continuous Measurement Control (CMC function) facilitates to evaluate
the current measured value instantly and definitely.
After each successful calibration the scale of the pH measuring range is
displayed in the measured value display. Here you can very clearly see
whether or not the current measured value is in the calibrated part of the
measuring range.
Error
Error
< -30 or
> 30
... -62 or
... -50
Perform error elimination according to
chapter 15 W
For pH IDS sensors, you can optionally enable a more finely
graded calibration evaluation (QSC) (see section 5.4).
Display
Calibration
record
Zero point
[mV]
Slope [mV/
pH]
Multi 3620 IDS
Ser. no. 09250023
CALIBRATIONpH
Calibration date
15.03.2016 16:13:33
SenTix 940
Ser. no. B092500013
TEC
Buffer 1
4.01
Buffer 2
7.00
Buffer 3
10.01
Voltage
1
184.0 mV 24.0 °C
Voltage 2
3.0 mV 24.0 °C
Voltage
3
-177.0 mV 24.0 °C
Slope
-60.2 mV/pH
Asymmetry
4.0 mV
Sensor
+++
etc...