Conductivity
Multi 3620 IDS
60
ba77169d06
10/2021
8.3.4 Calibration data
The calibration data can be displayed and then output to the interface.
Displaying the cali-
bration record
The calibration protocol of the last calibration is available in the menu
Calibra-
tion
/
Calibration record
. To open it in the measured value display, press the
<CAL_>
key.
The calibration records of the last 10 calibrations are available in the menu
Cali-
bration
/
Calibration data storage
/
Display
. To open the
Calibration
menu in the
measured value display, press the
<MENU/ENTER>
key.
3.
Start the calibration with
<CAL>
.
The cell constant that was calibrated last is displayed.
4.
Immerse the IDS conductivity sensor in the control standard solution,
0.01 mol/l KCI.
5.
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
6.
Wait for the measurement with stability control to be completed (status
indicator [HOLD][AR]) or
accept the calibration value with
<MENU/ENTER>
.
The calibration record is displayed and output to the interface.
7.
Switch to the measured value display with
<MENU/ENTER>
.
The calibration record is automatically transmitted to the interface
after calibrating.
30.09.2012 08:00