Settings
Multi 3620 IDS
76
ba77169d06
10/2021
ing the settings, switch to the measured value display with
<M>
.
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration record of the
last calibration.
Calibration
/
Calibration data
storage
/
Display
-
Displays the last calibration records
(max. 10)
Calibration
/
Calibration data
storage
/
Output to
USB flash drive or
printer
-
Outputs the stored calibration data to
the USB-A interface (
USB Host
) (USB
memory device/USB printer)
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the stored calibration data to
the USB-B interface (
USB Device
)
(PC)
Calibration
/
Calibration interval
1 ...
180
...
999 d
Calibration interval
for the D.O. sen-
sor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.
FDO Check
/
Start FDO Check
-
Starts the FDO
®
check procedure
FDO Check
/
Check interval
1 ...
60
...
999 d
Interval for the
FDO Check
(in days).
The meter reminds you to check the
sensor regularly by
FDO Check
status
indicator in the measuring screen.
Sal automatic
on
off
Automatic salt content correction for
concentration measurements. The
measured salinity value is taken over
by a connected conductivity sensor.
This menu item is only available if an
IDS conductivity sensor is connected.
Sal correction
on
off
Manual salt content correction for
concentration measurements.