Multi 3620 IDS
Settings
ba77169d06
10/2021
81
System
/
General
/
Brightness
0 ...
15
... 22
Changes the display brightness
System
/
General
/
Switchoff time
10 min ...
1h
... 24
h
Adjusts the switch-off time
System
/
General
/
Temperature
unit
°C
°F
Temperature unit,
degrees Celsius or degrees Fahrenheit.
All temperature values are displayed with
the selected unit.
System
/
General
/
Stability con-
trol
on
off
Switches on or off the automatic stability
control during measurement (see section
11.6.3)
System
/
Interface
/
Baud rate
1200, 2400,
4800
,
9600, 19200
Baud rate of the USB device interface
System
/
Interface
/
Output format
ASCII
CSV
Output format for data transmission
For details, see section 13
System
/
Interface
/
Decimal sep-
arator
Dot (xx.x)
Comma (xx,x)
Decimal separator
System
/
Interface
/
Output
header
Output of a header for
Output format
:
CSV
System
/
Interface
/
Extended Oxi
output
The measured values for the parameters
concentration (mg/l ) and saturation (%)
are output together.
The function is active if the following
requirements are met:
a D.O. sensor is connected
The D.O. sensor displays the parame-
ter concentration (mg/l) or saturation
(%)
the
Output format
CSV
is set
System
/
Clock function
Date format
Datum
Time
Settings of time and date.
For details, see section 4.5.5
System
/
Service
/
Meter informa-
tion
Hardware version and software version
of the meter are displayed.
System
/
Service
/
Reset of Oxi-
Top set management
Only for operation with OxiTop IDS mea-
suring heads.
Resets the OxiTop set management to
the default values.For details, see Com-
plementary operating manual
OxiTop
®
-IDS (/B) Multi 3620 IDS
wireless operation of OxiTop
®
-IDS (/B)
measuring heads
System
/
Reset
-
Resets the system settings to the default
values.
For details, see section 11.7.2
Menu item
Setting
Explanation