Data memory
Multi 3620 IDS
86
ba77169d06
10/2021
12 Data memory
You can transmit measured values (datasets) to the data memory:
Manual storage (see section 12.1)
Automatic storing at intervals (see section 12.2)
Each data storing process transmits the current dataset to the interface at the
same time.
12.1 Manual storage
You can store a measurement dataset to the data memory as follows. The
dataset is at the same time output to the interface:
If the memory is full
When all storage locations are occupied, it is not possible to continue storing.
Then you can e.g. transmit the stored data to a PC or a USB flash drive (see
section 12.3.1) and subsequently erase the storage (see section 12.3.2).
12.2 Automatic data storage at intervals
The storing interval (
Interval
) determines the time interval between automatic
data storing processes. Each data storing process transmits the current data-
set to the interface at the same time.
Configuring the
automatic memory
function
1.
Press the
<STO>
key shortly.
The menu for manual data storage appears.
2.
If necessary, change and confirm the ID number (1 ... 10000) with
<
><
>
and
<MENU/ENTER>
.
The dataset is stored. The meter switches to the measured value dis-
play.
15.03.2014 11:24:16
pH 7.000 24.8
°C AR +++
ID number
:
1
continue
Manual data storage 4 von 500
30.09.2012 08:00
1.
Press the
<STO_>
key.
The menu for automatic data storing appears.