17
Request
Function Code
1 byte
0x04
Register Address
2 bytes
0x0000
Number Registers
2 bytes
0x0020
Error
Error Code
1 byte
0x84
Exception Code
1 bytes
0x02 (Illegal Address)
Response
Function Code
1 byte
0x04
Byte Count
1 byte
0x40
Register Values
64 bytes
ODO Saturation (%) [3:0]
ODO (mg/L) [3:0]
ODO Local Barometer Compensated (%) [3:0]
Temperature (C) [3:0]
Ref Temperature (C) [3:0]
Time Since Boot (ms) [3:0]
Conductivity (us/cm) [3:0]
Specific Conductivity (us/cm) [3:0]
Salinity (ppt) [3:0]
Conductivity nLF (us/cm) [3:0]
Total Dissolved Solids (mg/L) [3:0]
Reserved [3:0]
Reserved [3:0]
Reserved [3:0]
Reserved [3:0]
Reserved [3:0]
Reserved [3:0]
Modbus Configuration & Commands