7.4
Settings
With this function you can undertake or change settings. The table
below shows the settings in the delivery state.
7.4.1
Operating mode
You can select the following operating modes:
BOD Standard
BOD Routine
GLP/TOOLS
menu.
Select the
- Settings
menu item with
<
><
>
GLP
Setting
Default setting
(bold)
and setting range
Operating mode
BOD Routine
BOD Standard
Operational life
5 days
(0.5 h to 99 days)
5 days
(0.5 h to 99 days)
Date
current
current
Time
current
current
GLP
Off
(permanently set)
On
Off
Calibration interval
---
12 months
to 36 months
Clearing the memory
automatic
or manual
automatic
or manual
AutoTemp
On
(permanently set)
On
Off
Switch-off interval
5 minutes
(permanently set)
5 minutes
to 15 minutes
Language
German
German
SETTINGS
menu:
Select the
- Operation mode
menu item with
<
><
>
.
RUN/ENTER