What to do if...
10.1
Display messages
Display message
Cause
Remedy
already used!
The selected measuring head has
already been started for a measure-
ment and can only be used for a new
measurement after the current mea-
surement is finished and the mea-
surement data completely retrieved.
Select a new and free measuring
head or
- Stop
start
or
Release the measuring head (sec-
tion 7.6.2 R
Please
Set system clock!
Data backup battery was changed.
Set the clock (see section 7.4.3
D
defective!
The selected measuring head is
defective.
Please return it to WTW.
Select a new and free measuring
head or
- Stop
start
No samples
available!
There are no finished samples stored
in the sample management.
Transfer the unfinished samples of
the sample management to “fin-
ished” status.
To do this, the measurement data
of all measuring heads must be
called up and evaluation and docu-
mentation of the measurement data
of the finished samples must be
performed.
If the measurement data of the
measuring heads are not yet fin-
ished, no instant remedy is possi-
ble!
The memory capacity limit of the
instrument has been reached.
Summary of Contents for OxiTop Control OC 100
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents OxiTop Control OC 100 6 ba31114e07 08 2014...
Page 14: ...Commissioning OxiTop Control OC 100 14 ba31114e07 08 2014...
Page 20: ...Operation OxiTop Control OC 100 20 ba31114e07 08 2014...
Page 82: ...OxiTop Control OC 100 82 ba31114d07 08 2014...
Page 83: ......