OxiTop
®
Control OC 100
What to do if...
ba31114e07
08/2014
71
Memory lack!
No ready samples
autom. erasable!
The instrument works with the set-
ting, Erase memory
auto
(automati-
cally). The memory is full and only
unfinished samples are stored in the
sample management.
The instrument can only erase fin-
ished samples automatically.
Transfer the unfinished samples of
the sample management to “fin-
ished” status.
To do this, the measurement data
of all measuring heads must be
called up and evaluation and docu-
mentation of the measurement data
of the finished samples must be
performed.
If the measurement data of the
measuring heads are not yet fin-
ished, no instant remedy is possi-
ble!
The memory capacity limit of the
instrument has been reached.
Memory lack!
Erase finished
sample/s!
The instrument works with the set-
ting, Erase memory "manually". The
memory is full.
Transfer the unfinished samples of
the sample management to “fin-
ished” status:
Call up the measurement data of all
measuring heads
Perform the evaluation and docu-
mentation for the measurement
data of the finished samples
Erase finished samples from the
sample management:
section 7.6.1 E
Note: Only the finished samples in the
sample management can be erased.
Display message
Cause
Remedy
Summary of Contents for OxiTop Control OC 100
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents OxiTop Control OC 100 6 ba31114e07 08 2014...
Page 14: ...Commissioning OxiTop Control OC 100 14 ba31114e07 08 2014...
Page 20: ...Operation OxiTop Control OC 100 20 ba31114e07 08 2014...
Page 82: ...OxiTop Control OC 100 82 ba31114d07 08 2014...
Page 83: ......