OxiTop
®
Control OC 100
What to do if...
ba31114e07
08/2014
77
Check of the controller
Controller information
All single information on the state of the controller is listed. The check can
be undertaken at any time without affecting the measurements that are run-
ning.
Perform the command:
- All
. See
GLP/TOOLS
- Check
- Show
.
(The check is used in this case to test the IR interface.) All working measur-
ing heads must flash for 5 seconds.
Perform a measuring head check (see section 7.5.2 M
, page 51). (The check is used in this case to test the IR inter-
face.) Here, a working measuring head should supply its status data.
Keyboard, display and signal tone should show the required reaction.
Check the function of the clock (prerequisite: Clock is set).
Switch off the instrument and switch it on again. Time and date each appear
showing the current values.
Check RS232 interface (only required when using the WTW software,
ACHAT OC):
– Connect the controller to your PC by means of the interface cable
AK540/B.
– Switch on the controller and start the PC program, ACHAT OC. In doing
so, the PC program checks the RS232 interface.
The controller does not
react / is defective – the
data of the running mea-
surements should be
restored
Send the controller to WTW. Using a repaired or new controller (memory must
be empty!) perform the “Restore data” procedure (see section 7.6 M
Incorrect time display on
being switched on
Data backup battery was changed.
Summer/winter time change has
taken place.
Set the clock (see section 7.4.3
D
Cause
Remedy
Summary of Contents for OxiTop Control OC 100
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents OxiTop Control OC 100 6 ba31114e07 08 2014...
Page 14: ...Commissioning OxiTop Control OC 100 14 ba31114e07 08 2014...
Page 20: ...Operation OxiTop Control OC 100 20 ba31114e07 08 2014...
Page 82: ...OxiTop Control OC 100 82 ba31114d07 08 2014...
Page 83: ......