OxiTop
®
Control OC 100
Operating mode: BOD Standard / BOD Routine
ba31114e07
08/2014
21
5
Operating mode:
BOD Standard
/
BOD Rou-
tine
The
BOD Standard
mode (parallel sample process for up to 12 measuring
heads per overall sample) is preset in the delivery condition of the controller.
To change to the
BOD Routine
operating mode (single samples): see section
The handling of controller and measuring heads is basically the same in both
operating modes. Where there are differences for the
BOD Routine
operating
mode they are described.
5.1
Sample preparation
See application reports.
Screw the OxiTop
®
measuring heads onto the measuring vessels and
close them tightly.
NOTE
Never use joint grease or other lubricants for your OxiTop
®
measuring
heads. Some of these products contain solvents that can cause severe
damage to the plastic housing of the measuring head.
The sealing of the measuring vessels is perfectly adequate without
grease. However, you should always wipe off heavy contamination and
particles on the sealing surfaces of the rubber sleeves and OxiTop
®
measuring heads. WTW accepts no liability for damage due to the use of
joint grease.
5.2
Starting the measurement
SELECT ACTION
menu (com-
munication with the measuring
heads):
Preselected:
- Start sample
.
(In
BOD Routine
mode appears:
START
).
Select the measuring range with
<
><
>
.
The filling volume required is
given in the right hand column.
The controller stores the setting
(memory function: the last
selected measurement range is
set).
RUN/ENTER
Summary of Contents for OxiTop Control OC 100
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents OxiTop Control OC 100 6 ba31114e07 08 2014...
Page 14: ...Commissioning OxiTop Control OC 100 14 ba31114e07 08 2014...
Page 20: ...Operation OxiTop Control OC 100 20 ba31114e07 08 2014...
Page 82: ...OxiTop Control OC 100 82 ba31114d07 08 2014...
Page 83: ......