GLP/TOOLS
OxiTop
®
Control OC 100
60
ba31114e07
08/2014
7.6.3
Reading out data
In case you lost your controller or your controller is defective, the
"Restore data" function facilitates to restore the data of running
measurements using a new or other controller. This means the data are
not lost!
To perform this function the memory of the controller must be abso-
lutely empty! The data of each single measuring head are restored
one after the other in a sequence. It is not possible to leave this
function and to continue it afterwards because then the memory
would no longer be empty.
Select the
- Restore data
menu
item.
Select the
- Restore
data
menu item.
Confirm the restoration for one
measuring head and hold the
controller to the measuring
head. Contact selection:
The controller displays the
serial number of the selected
measuring head.
After confirming again the data
of the measuring head are read
out and stored in the controller.
RUN/ENTER
RUN/ENTER
RUN/ENTER
The Id number of the sample is not restored, it is always
I
999.