13 / 25
Menu structure
6.2
7
Data communication / printer
PC-Software „STMLabReader“
7.1
The reading of the data logger as well as the configuration of the measuring system can also take place via the
software „STMLabReader“.
Use only interface cable (Interface HMG USB) to transfer data from the meter to the PC. This cable is not
included in the delivery and must be purchased separately (
chapter „Accessories”).
To use this function, press (
CONF
). This will take you to the menu item „
Configuration
“ and on to the
submenus item „
protocol
“ choose
„standard”.
You will find the software here: http://download.meinsberg.de/en/
Serial data output via ASCII
7.2
The current measured values can be cyclically output via the serial interface as ASCII characters.
Use only interface cable (Interface HMG USB) to transfer data from the meter to the PC. This cable is not
included in the delivery and must be purchased separately (
chapter „Accessories”).
To use this function, press (
CONF
). This will take you to the menu item „
Configuration
“ and on to the
submenus item „
protocol
“ choose
„ASCII”
and setup the
„cycle time”.
The interface setting of the receiver is 4800 bauds, 1 stop bit and no parity. All output characters are in the
ASCII format.
Data decription: dd mm yy hh minmin E1E1 MMMM E2E2 NNNN E3E3 TTTT
Dd
2 Byte
Day
mm
2 Byte
Month
yy
2 Byte
Year
hh
2 Byte
Hour
minmin
2 Byte
Minute
E1E1
4 Byte
Unit
MMMM
4 Byte
measured main value
E2E2
4 Byte
Unit
NNNN
4 Byte
Secondary measured value
E3E3
4 Byte
Unit
TTTT
4 Byte
temperature
MIN/MAX INFO
RESET MIN/MAX
MESSMODUS
CAL INFORMATION