23 / 25
This completes the calibration process. The new calibration value will be saved in the device.
9.4.2.4 Automatic calibration
The automatic calibration of the conductivity sensor is a one point calibration with automatic buffer detection.
The following conductivity calibrations are archived:
0.01 N KCl:
1.41 mS/cm (25 °C)
0.1 N KCl:
12.9 mS/cm (25 °C)
> You will be prompted to immerse the sensor into the calibration medium.
Confirm this dialog with (
ON/OFF
). The currently measured values will be displayed.
If the measured values (conduc temperature!) are in a stable condition,
reconfirm by pressing (
ON/OFF
).
This completes the calibration process. The new calibration value will be saved in the device.
If the calibration is faulty, an error message will be displayed (see
chapter 9.4.2.5).
9.4.2.5 Calibration error
The limits for all calibration values are saved in the meter. If the entered values are outside these limits, one of
the following error messages will be displayed:
„!
Incorrect entry!
“ or „
!Calibration error!
“.
You may have to repeat the calibration, service the sensor or replace it (wear part!).
10
Scope of delivery
PTM40/Set
Lab pH meter
Power supply, Tripod
pH Combination Electrode EGA142-K010-U-X
2x 50 ml buffer solution pH 4.01; 6.86 (25 °C)
PLF40/Set
Lab conductivity meter
Power supply, Tripod
Conductivity Cell LTC0,35/23-K010-U
50 ml calibration solution 0.1 N KCl (12.9 mS/cm at 25 °C)
11
Accessories
Tripod set TMG
adapter, rod and electrode holder for PTM40/PLF40
Power supply TMG
extended range power supply for PTM40/PLF40
Interface HMG USB
USB cable with data transfer software for PTM40/PLF40
Interface TMG Printer
USB cable with data transfer to a printer for PTM40/PLF40
pH Combination Electrode
plastic shaft, gel electrolyte, 2 ceramic frits, built-in temperature sensor
EGA142-K010-U-X
Pt 1000, 1 m fixed cable with 6 pole connector BK, length 120 mm,
Ø 12 mm, -5...+80 °C
Conductivity Measuring Cell
plastic shaft, sensor material graphite, cell constant 0.33 cm
-1
LTC0,35/23-K010-U
4-electrode technology, built-in temperature sensor Pt 1000,
1 m fixed cable with 6 pole connector BK, length 120 mm,