background image

User Manual 
Modbus RTU 
Option module for PolluTherm

 

 

M H 6123 INT, page 3 

4.  Modbus data register list 

Designation Modbus 

Register 

Modbus 

Register Type 

Modbus 

Address 

Data Value 

Range 

Unit

Read only (RO) 

Read/write (R/W) 

Firmware 
Version

2

 

30001 or 40001  Input or Holding  0 

Uint16 

RO 

Energy 

30002 or 40002  Input or Holding  1 

Int32 

kWh, MWh, 
MJ, GJ 

RO 

Energy  
(Unit factor) 

30004 or 40004  Input or Holding  3 

UInt16 

RO 

Energy (Unit) 

30005 or 40005  Input or Holding  4 

4 char ASCII 

RO 

Energy (Float) 

30007 or 40007  Input or Holding  6 

IEEE754 

RO 

Volume 

30009 or 40009  Input or Holding  8 

Int32 

l (dm

3

), m

3

 

RO 

Volume  
(Unit factor) 

30011 or 40011  Input or Holding  10 

UInt16 

RO 

Volume (Unit) 

30012 or 40012  Input or Holding  11 

4 char ASCII 

RO 

Volume (Float) 

30014 or 40014  Input or Holding  13 

IEEE754 

RO 

Flow 

30016 or 40016  Input or Holding  15 

Int32 

l/h, m

3

/h 

RO 

Flow  
(Unit factor) 

30018 or 40018  Input or Holding  17 

UInt16 

RO 

Flow (Unit) 

30019 or 40019  Input or Holding  18 

4 char ASCII 

RO 

Flow (Float) 

30021 or 40021  Input or Holding  20 

IEEE754 

RO 

Power 

30023 or 40023  Input or Holding  22 

Int32 

W, kW 

RO 

Power  
(Unit factor) 

30025 or 40025  Input or Holding  24 

UInt16 

RO 

Power (Unit) 

30026 or 40026  Input or Holding  25 

4 char ASCII 

RO 

Power (Float) 

30028 or 40028  Input or Holding  27 

IEEE754 

RO 

Inlet 
temperature 
(Fixed) 

30030 or 40030  Input or Holding  29 

Int16 

0.01°C 

RO 

Inlet 
temperature 
(Float) 

30031 or 40031  Input or Holding  30 

IEEE754 

°C 

RO 

Outlet 
temperature 
(Fixed) 

30033 or 40033  Input or Holding  32 

Int16 

0.01°C 

RO 

Outlet 
temperature 
(Float) 

30034 or 40034  Input or Holding  33 

IEEE754 

°C 

RO 

Temperature 
difference 
(Fixed) 

30036 or 40036  Input or Holding  35 

Int32 

0.001°C 

RO 

Temperature 
difference 
(Float) 

30038 or 40038  Input or Holding  37 

IEEE754 

°C 

RO 

Tariff Energy 

30040 or 40040  Input or Holding  39 

Int32 

kWh, MWh, 
MJ, GJ 

RO 

Tariff Energy 
(Unit factor) 

30042 or 40042  Input or Holding  41 

UInt16 

RO 

Tariff Energy 
(Unit) 

30043 or 40043  Input or Holding  42 

4 char ASCII 

RO 

Tariff Energy 
(Float) 

30045 or 40045  Input or Holding  44 

IEEE754 

RO 

Cooling Energy  30047 or 40047  Input or Holding  46 

Int32 

kWh, MWh, 
MJ, GJ 

RO 

Cooling Energy 
(Unit factor) 

30049 or 40049  Input or Holding  48 

UInt16 

RO 

Cooling Energy 
(Unit) 

30050 or 40050  Input or Holding  49 

4 char ASCII 

RO 

Cooling Energy 
(Float) 

30052 or 40052  Input or Holding  51 

IEEE754 

RO 

Fabrication 
Number 

30054 or 40054  Input or Holding  53 

Int32 

RO 

Reviews: