HandyLab 100
Operation
25
ba77062d02
12/2017
Stability control
(AutoRead )
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic
Stability control
is switched off.
You can start the
Stability control
manually at any time, irrespective of
the setting for automatic
Stability control
(see page 23) in the
Measure-
ment
menu.
3
Select the pH or mV display with
<M>
.
pH
6.949
25.0
°C
15.03.2014
08:00
Menu
1
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
The HOLD function is active.
2
Using
<ENTER>
, activate the
Stability control
function manu-
ally.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized.
Note
You can prematurely terminate the
Stability control
function manually
with
<ENTER>
at any time. If the
Stability control
function is prema-
turely terminated, the current measurement data are output to the
interface without the AutoRead info.