HandyLab 100
Operation
23
ba77062d02
12/2017
4.3.2
Data storage
This menu contains all functions to display, edit and erase stored mea-
sured values.
Note
Detailed information on the memory functions of the HandyLab 100 is
given in section 4.5.
4.3.3
Automatic
Stability control
The automatic
Stability control
(AutoRead) function continuously
checks the stability of the measurement signal. The stability has a con-
siderable impact on the reproducibility of measured values.
You can activate or switch off the automatic
Stability control
function
(see section 4.4.4).
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when you switch over between the measured parameters with
<M>
.
when the automatic
Stability control
is switched off.
4.4
pH value / ORP voltage
4.4.1
General information
You can measure the following parameters:
pH value [ ]
ORP [mV]
Temperature
measurement
For reproducible pH measurements, it is essential to measure the tem-
perature of the test sample.
You have the following options to measure the temperature:
Automatic measurement of the temperature by the temperature sen-
System
/
Reset
-
Resets the system settings
to the default values.
For details, see section
4.6.2
Menu item
Setting
Explanation