Lab 865
Settings
ba77145e02
12/2017
37
7
Settings
7.1
Measurement settings
7.1.1 Changing the settings for pH measurements
The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required parame-
ter in the measured value display and press the
<F1>
/[menu] or
<ENTER>
key.
After completing the settings, switch to the measured value display with
<M>
.
Default settings are printed in
bold
.
Menu item
Possible set-
ting
Description
Calibration
/
Cali-
bration record
-
Displays the calibration record of
the last calibration.
Calibration
/
Calibration data
storage
/
Display
-
Displays the calibration record.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibration
record to the interface with
<F2>
/
[USB output]
.
Output all calibration records to
the interface with
<F2__>
[USB
output]
.
Quit the display with
<F1>
/
[Back]
or
<ENTER>
.
Switch directly to the measured
value display with
<M>
.
Calibration
/
Cali-
bration data stor-
age
/
Output to USB
-
Outputs the calibration records to
the interface.
Calibration
/
Serial
number (probe)
Entry of the serial number of the
connected sensor.
The serial number is output in the
calibration record.
Change the content of the high-
lighted position with
<
><
>
.
Go to the next position with
<F2>
/[
].
When the serial number has
been completely entered, con-
firm with
<ENTER>
.