3.2 Selection of the viscometer
The flow time of the liquid to be measured depends on the size of the capillaries. The capillary size of
the viscometer should be selected in such a manner that flow times greater than 100 s result
.
We only guarantee original SI Analytics
®
viscometers that they will fit into the ViscoClock
plus
and that the
calibration of the calibrated viscometer models yields correct measured values.
Viscometer
Type No.
uncalibrated
Calibrated for measurements
Viscosity determination
manual
automatic
relative
absolute
DIN-Ubbelohde
532…
-
-
DIN-Ubbelohde
501…
-
-
-
DIN-Ubbelohde
530…
-
-
-
ASTM-Ubbelohde
527…
-
-
ASTM-Ubbelohde
525…
-
-
-
ASTM-Ubbelohde
526…
-
-
-
Micro Ubbelohde
537…
-
-
Micro Ubbelohde
536…
-
-
-
Micro Ubbelohde
538…
-
-
-
Micro Ostwald
517…
-
-
Micro Ostwald
516…
-
-
-
Micro Ostwald
518…
-
-
-
Table 1 List of viscometers for the selection of viscometers that are suitable to measure the flow
time for the application and to determine the viscosity by means of the ViscoClock plus.
Based on their construction, viscometers type 502 41, type 502 43 and type 502 50 are not suitable!
To measure the absolute kinematic viscosity, you
must
use viscometers that have been calibrated for
automatic measurements!
Prior to the first use, we recommend a cleaning procedure as per DIN 51 562, part 1.
After this, the viscometer must be dried and free of dust!
1
To measure the absolute (kinematic) viscosity with Ubbelohde viscometers, many standards assume
minimum run times of 200 seconds. Shorter run times are permitted if you use Micro Ubbelohde
viscometers or for measuring relative viscosities.