5.11.3 Decommissioning the measuring unit
figure 5-23Tab Wartung (example Alyza IQ PO4)
4
Carry out the function
Wartungszustand einschalten
.
On the IQ S
ENSOR
N
ET
, the maintenance condition for the Alyza IQ
is
switched on.
5
Carry out the
Alyza IQ STOPPEN
function to stop the running operation.
1
Open the
Wartung Messeinheit
menu.
2
Select the function,
Öffnen der Messeinheit vorbereiten
.
Follow the instructions on the display.
The procedure starts the temperature adjustment of the measuring unit.
3
Drain the tubes (
Wartung
/
Manuelle Funktionen
/
System entleeren
).
4
Open the outer housing door far enough so the arrestable brake-stay
(on the lower right side of the housing) catches.
5
Open the inner housing door far enough so that it touches the outer
door. Tilt the angular sheet (at the inside of the outer door) over the
upper edge of the inside door.
The measuring unit is ready to be opened when the status LED of the
measuring unit lights up white.