Alyza IQ PO4
Overview
15
ba76201d03
05/2020
agents, standard solutions, cleaning solution).
The MultiPort Valve (4) runs each of the moving liquids to the place where they
are required. The dosing of the reagents to the sample takes place in the mixing
chamber in the MultiPort valve.
The sample mixed with reagents is then moved to the cell in the photometer unit
(9) to be measured.
14 shows the open photometer unit (without front cover).
The photometer unit has an LED as the light source and a photo diode as the
detector. After the measurement, the liquid is removed from the photometer unit
through the waste tube (4).
1.3.3 ChemBags
The Alyza IQ has an extra counter for each liquid container. The counter counts
the consumption of the following procedures as soon as the function is started:
Measuring
Calibration
Cleaning
figure 1-6 Photometer unit (without front cover)
1
Photometer
2
Cell holder with connectors for tubes
3
Photometer feed tube (MPV mixing chamber - cell)
4
Photometer drain tube (cell - waste tube)
1
3
4
2