Commissioning
Alyza IQ PO4
28
ba76201d03
05/2020
(according to the technical data of the instrument input or output)
– Overvoltage class II surge limiters
Suitable disconnecting device (e.g. switch or circuit breaker) for the line
power supply of permanently mounted equipment with separate line power
connection,
– labeled as disconnecting device for this instrument
– compliant with the following regulations
- IEC 60947-1
- IEC 60947-3
– in the vicinity of the instruments (recommendation)
Fault current protection switch (ground fault circuit interrupter)
especially with operation of heat tracings
Flame resistant (cable and lines),
compliant with the following regulations
– UL 2556 VW-1 (for USA, Canada)
– IEC 60332-1-2 (outside of USA, Canada)
3.3.3 General installation instructions
Pay attention to the following points during installation:
Due to its weight, the Alyza IQ always has to be carried by two people (hous-
ing door upward, both people grasping the housing at the upper C rail and at
the housing bottom on the side of the door).
Wear safety shoes for transport, installation and mounting work (see
section 2.4 Personal protective equipment (PPE),
Installation in the open should be done while the weather is frost-free
(environmental conditions, see section 3.3.1 Requirements of the measure-
ment location,
Mount the Alyza IQ as straight as possible to ensure that the liquids can drain
off optimally.
Mount the Alyza IQ so that the space under the housing bottom is always free
for ventilation of the housing.
Details on the conditions at the installation site:
See chapter 7 Technical data,
This section describes the installation of the Alyza IQ with various,
especially designed accessories. We assume that the operator uses
these accessories. In this section, the individual scopes of delivery
are not distinguished so the comprehensibility of the operating man-
ual is not affected.