Commissioning
Alyza IQ PO4
46
ba76201d03
05/2020
Installation
figure 3-17Mounting the bug screen and condensate drain adapter
1
Condensate drain tube (black) at the housing bottom
2
Bug screen
3
Frame with seal
4
Condensate drain adapter with 2 screws
5
Condensate drain tube (transparent)
6
4 knurled-head screws
1
2
3
4
5
6
1
Screw the condensate drain adapter (4) with 2 screws to the frame (3)
so that the tube nozzle of the adapter is on the outside of the frame.
2
Plug the transparent condensate drain tube (5) onto the tube nozzle of
the condensate drain adapter (4) as far as it will go.
3
Insert the bug screen (2) in the frame so that it is kept in position by the
seal of the frame (3).
4
Position the frame (3) with the bug screen (2) and the mounted conden-
sate drain adapter (4) at the underside of the housing. When doing so,
the black condensate drain tube (1) at the underside of the housing
should exactly fit into the recess of the condensate drain adapter (4).
5
Screw the frame (4) to the housing bottom using the 4 knurled-head
screws.