Alyza IQ PO4
Measurement / Operation
97
ba76201d03
05/2020
4.6.1 Transferring to a USB memory device a selection of important oper-
ating data
These operating data provide an overview of important settings and data for the
functioning of your Alyza IQ.
Measurement settings
Calibration settings
Calibration data of the last calibration
Lifetimes of the replacement parts (MPV, ChemBags)
4.6.2 Transferring detailed operating data to a USB memory device for
evaluation by the service department
The service department can make use of the detailed operating data to analyze
and eliminate errors.
1
Plug a USB memory device to the USB interface of your IQ S
ENSOR
N
ET
controller (e.g. Terminal/Controller 2020 3G).
2
At the terminal, open the tab
Wartung
in the Alyza menu.
3
Select the function,
Save Alyza info file to USB memory device
.
Follow the instructions on the display.
4
Remove the USB memory device from the USB interface of your
IQ S
ENSOR
N
ET
controller.
5
Connect the USB memory device to the USB interface of a PC and dis-
play the file system of the USB memory device.
On the USB memory device there is the folder "AlyzaInfoData" with the
info data of the Alyza IQ.
Each time the data are output a new file with the creation date and cre-
ation time in the file name is saved to the folder.
The assignment of the data to an Alyza IQ is secured via the series
number of the Alyza IQ in the file name.
1
Unscrew the 2 fixing screws of the cover lid of the control unit ACM and
remove the cover lid of the ACM.
2
If a USB plug is connected to the USB interface next to the "USB0" inter-
face:
Remove this USB plug.
3
Plug a USB memory device to the USB interface "USB0".
4
At the terminal, open the tab
Wartung
in the Alyza menu.
Summary of Contents for WTW Alyza IQ PO4
Page 171: ......