Technical data
Alyza IQ PO4
162
ba76201d03
05/2020
Ambient
conditions
Components
Temperature range
Mounting/installation/
maintenance
+ 5 ... + 40 °C (+ 41 ... +104 °F)
Operation
- 20 … + 40 °C ( - 4 … + 104 °F)
Storage
- 20 …+ 50 °C (- 4 … + 122 °F)
(in completely empty condition)
Relative air humidity
Max. 95 % (noncondensing)
Mounting/installation/
maintenance
≤
80 %
Yearly average
≤
90 %
Dew formation
Possible
Measuring location
Indoors and outdoors
Site altitude
Max. 4000 m above sea level
Level of contamination
2
Chloride concentration
< 500 mg/l (no sea water)
Housing
Sunlight (UV) resistant
Material: powder-coated aluminium
Mounting plate: PVC
Cable glands at the base
plate
M40 x 1.5 (black, large):
Clamping range 19 - 28 mm
M20 x 1.5 (black, small):
Clamping range 6.5 - 12 mm
Overflow vessel
Material: PMMA
Required sample quantity:
1... 25 ml/min
For correct measurements, enough sample
must always be available in the overflow ves-
sel.
Temperature control
Heating, cooling, ventilation
Light source for photome-
ter
LED, 400 nm (violet)
MultiPort valve
(MPV)
Materials: PMMA, fluoroplastic, aluminum