Alyza IQ PO4
Technical data
163
ba76201d03
05/2020
Accessories
Meter safety
Test certificates
CE, cETLus
7.4
Electrical data
Mounting stand, wall
mount and rail mount
Rails: Stainless steel V2A,
Screws etc.: stainless steel V2A, V4A
Filtration unit
Membrane surface of the filter plate: 1300 cm
2
Cut-off limit: < 0.45 µm
Max. operating temperature: 45 °C (113 °F)
Max. suction height: 5 m
pH value: 2…11,5
Housing: PVC
Sleeve tube: PVC-reinforced PCV tube, trans-
parent, 24 x 3 mm
Intake line: PE 4 x 1 mm
Line length (intake line, return line): Max. 20 m
Applicable norms
EN 61010-1
UL 61010-1
CAN/CSA C22.2#61010-1
Electromagnetic compati-
bility
EN 61326-1, EN 61326-2-3,
FCC 47 CFR Part 15
Type of protection (hous-
ing)
IP 54 (EN 60529)
Type of protection
(internal power supply box)
IP 67 (EN 60529)
Protective class
I
Overvoltage category
II
Input voltage
120 V / AC ±10% / 60 Hz
Fuse protection of the
building
16 A with ground fault circuit interrupter
Line power connection
3-pin, N/L /PE
Line cross-section of line
power connection
Europe: 1.5 ... 4.0 mm
2
USA: AWG 14 ... 12
Summary of Contents for WTW Alyza IQ PO4