Cond 7110
Conductivity
ba75927e05
12/2017
23
Calibration evaluation (conductivity)
After calibration, the meter automatically evaluates the current status of the
calibration.
The calibration evaluation is displayed as a sensor symbol.
1/cm
9
0
7
4
C
Cal
Display
Cell constant [cm
-1
]
Within the range
0.450 ... 0.500 cm
-1
or
0.800 ... 0.880 cm
-1
E3
Outside the range
0.450 ... 0.500 cm
-1
or
0.800 ... 0.880 cm
-1
Eliminate the error
(see section 9 W
The calibration evaluation on the display (sensor symbol) flashes
when the set cleaning interval has expired and thus reminds you
to regularly clean and calibrate the measuring cell (see section
6.1.2 C
Summary of Contents for wtw Cond 7110
Page 1: ...OPERATING MANUAL ba75927e05 12 2017 Cond 7110 CONDUCTIVITY METER...
Page 2: ...Cond 7110 2 ba75927e05 12 2017 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents Cond 7110 6 ba75927e05 12 2017...
Page 42: ...Index Cond 7110 42 ba75927e05 12 2017...
Page 43: ......