Settings
Cond 7110
24
ba75927e05
12/2017
6
Settings
The meter has separate setting routines for the measurement settings and
system settings.
6.1
Measurement settings (conductivity)
6.1.1
Changing the settings for conductivity measurements
1.
Open the setting menu in the measured value display with
<MODE
_
>
.
The first setting is displayed.
2.
If necessary, indicate the required setting with
<ENTER>
.
3.
To change the current setting, press
<
><
>.
4.
Confirm the setting with
<ENTER>
.
The setting is stored.
The next setting is displayed.
5.
Change or confirm the other settings.
or
Quit the setting menu with
<MODE>
.
List of settings for conductivity measurements
Default settings are printed in
bold
.
The settings appear in the following order:
When the last setting is confirmed, the setting menu is auto-
matically quit.
Displayed
(Confirm with
<ENTER>)
Possible setting
(Change with
<
><
>)
Description
t25 /t20
TRef25
Tref20
Reference temperature
Uni
°C
°F
Temperature unit
Int.C
1 ...
180
... 999 d
Cleaning interval
Summary of Contents for wtw Cond 7110
Page 1: ...OPERATING MANUAL ba75927e05 12 2017 Cond 7110 CONDUCTIVITY METER...
Page 2: ...Cond 7110 2 ba75927e05 12 2017 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents Cond 7110 6 ba75927e05 12 2017...
Page 42: ...Index Cond 7110 42 ba75927e05 12 2017...
Page 43: ......