Technical data
Cond 7110
34
ba75927e05
12/2017
10
Technical data
10.1
Measuring ranges, resolution, accuracy
10.1.1
Measuring ranges, resolution
Parameter
Measuring range
Resolution
ϰ
[µS/cm]
0.000 ... 1.999
*
0.00 ... 19.99
**
0.0 ... 199.9
200 ... 1999
* only possible with cells of the cell constant, 0.010 cm
-1
** only possible with cells of the cell constant, 0.010 cm
-1
or 0.090 ... 0.110 cm
-1
0.001
0.01
0.1
1
ϰ
[mS/cm]
2.00 ... 19.99
20.0 ... 199.9
200 ... 1000
0.01
0.1
1
Specific resis-
tance [MOhm*cm]
1.000 ... 1.999
2.00 ... 19.99
20.0 ... 199.9
200 ... 1999
0.001
0.01
0.1
1
SAL
0.0 ... 70.0
according to the IOT table
0.1
TDS [mg/l]
0 ... 1999
Factor can be set between
0.40 and 1.00
1
T [°C]
- 25.0 ... + 125.0
0.1
T [°F]
- 13.0 ... + 257.0
0.1