Ion concentration
Multi 3320
42
ba77053e04
07/2018
Continuing with two-
point calibration
ISE
0.0
24.8
°C
Ag
0.020
mg/l
2
15.02.2014
12:12
Standard
mV
12.
Thoroughly rinse the electrode with distilled water.
13.
Immerse the electrode in standard solution 2.
14.
When calibrating without temperature sensor:
Measure the temperature of the standard solution using a thermom-
eter.
Use
<F2>
/[
⇅
] to select the setting of the temperature.
Use
<
> <
>
to set the temperature.
Use
<F2>
/[
⇅
] to select the setting of the concentration.
15.
Set the concentration of the standard solution with
<
> <
>
and
press
<ENTER>
.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
16.
Wait for the end of the AutoRead measurement or accept the calibra-
tion value with
<ENTER>
.
The calibration display for the next standard solution appears.
ISE
8.9
24.8
°C
15.02.2014
12:12
0.02 mg/l
mV
AR
ISE
20.5
24.8
°C
Ag
0.050
mg/l
3
15.02.2014
12:12
Standard
mV