Multi 3320
Dissolved oxygen
ba77053e04
07/2018
49
4.
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
5.
Wait for the end of the measurement with stability control or accept the
calibration value with
<ENTER>
.
The factor that was set last is displayed.
6.
Using
<
> <
>
, set the correction factor to adjust the displayed con-
centration value to the nominal value (value of the comparison mea-
surement). Subsequently, accept the correction factor with
<ENTER>
.
The meter switches to the measured value display.
The status display
[Factor]
is active.
O
2
7.92
mg/l
0.973
OxiComp
15.02.2014
08:00