Multi 3320
Conductivity
ba77053e04
07/2018
55
9.2.4
Calibration data
Displays the calibra-
tion data
The calibration record of the last calibration is to be found under the menu item,
<ENTER>
/
Calibration
/
Calibration record
. To open it, press the
<CAL
_
>
key
in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e.g. to a PC, with the
<F2>
[USB output]
key.
Displaying the cali-
bration data memory
The calibration records of the last calibration procedures are available in the
<ENTER>
/
Calibration
/
Calibration data storage
and
<ENTER>
/
Storage &
config
/
Data storage
/
Calibration data storage
menu.
4.
Immerse the conductivity measuring cell in the control standard solu-
tion, 0.01 mol/l KCI.
5.
Start the measurement with
<ENTER>
.
The [AR] status indicator is displayed. A progress bar is displayed and
the display of the measured parameter flashes.
6.
Wait for the end of the measurement with stability control or accept the
calibration value with
<ENTER>
.
The calibration record is displayed and output to the interface.
7.
Using
<F1>
/
[Continue]
or
<ENTER>
, switch to the measured value
display.
The calibration record is automatically transmitted to the interface
after calibrating.
ϰ
0.479 1
/cm
25.1
°C
15.02.2014
08:00