Multi 3320
Settings
ba77053e04
07/2018
63
ISE setup
/
AutoRead criterion
low
medium
high
Selection of the AutoRead
criteria (see section 7.1.1
M
ISE setup
/
Ion type
Ag, Br, Ca, Cd, Cl,
CN, Cu, F, I, K, Na,
NO3, Pb, S, NH3,
NH4*, CO2, ION
Selection of the ion type to
be measured.
An ion that is not included
in the list can be measured
with the setting, ION.
* Measuring with the NH 500 electrode:
The NH4 setting is not suitable for the gas-sensi-
tive electrode NH 500.
Select the following settings:
Ion type
"ION",
Valency
"-1".
ISE setup
/
Unit
mg/l
µmol/l
mg/kg
ppm
%
Selection, with which unit
the measurement result
and calibration standards
should be displayed.
ISE setup
/
Ion type
/ION
Valency
Molar mass
Set the valence (
Valency
)
and molar weight (
Molar
mass
) for the ion.
ISE setup
/
Density
0.001 ... 9.999 g/ml
or kg/l
Adjustable density of the
test sample (only with
Unit
:
mg/kg, ppm, %)
Menu item
Possible setting
Explanation