Settings
Multi 3320
64
ba77053e04
07/2018
10.1.6 Settings for D.O. sensors
The settings are available in the menu for measurement and calibration set-
tings. To open the settings, display the required measured parameter in the
measured value display and press the
<ENTER>
key. After completing the set-
tings, switch to the measured value display with
<M>
.
Default settings are printed in
bold
.
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration record of
the last calibration.
Calibration
/
Calibration data
storage
/
Display
-
Displays the calibration record.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibration
record to the interface with
<F2>
/
[USB output]
.
Output all calibration records to
the interface with
<F2
_
>
/
[USB
output]
.
Quit the display with
<F1>
/
[Back]
or
<ENTER>
.
Switch directly to the measured
value display with
<M>
.
Calibration
/
Calibration data
storage
/
Output to
USB
-
Outputs the calibration records to
the interface.
Calibration
/
Calibration interval
1 ...
14
...
999 d
Calibration interval
for the D.O. sen-
sor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring screen.
Calibration
/
Comparison meas.
On
Off
Enables to adjust the measured
value with the aid of a comparison
measurement, e.g. Winkler titration.
For details, see section 8.2 C