Multi 3320
Settings
ba77053e04
07/2018
65
10.1.7 Calibration interval
The calibration evaluation is displayed as a sensor symbol.
The sensor symbol flashes after the adjusted calibration interval has expired. It
is still possible to measure.
Setting the calibra-
tion interval
The calibration interval is set to 14 days in the factory.
You can change the interval (1 ... 999 days):
Sal correction
On
Off
Manual salt content correction for
concentration measurements.
Salinity
0.0
... 70.0
Salinity or salinity equivalent for the
salt content correction.
This function is only available for
concentration measurements if the
manual salt content correction is
switched on.
Reset
-
Resets all sensor settings to the
delivery condition (see section
10.3.1 R
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
Menu item
Possible
setting
Explanation
1.
Open the menu for measurement settings with
<ENTER>
.
2.
In the
Calibration
/
Calibration interval
menu, set the calibration interval
with
<
><
>
.
3.
Confirm the setting with
<ENTER>
.
4.
Quit the menu with
<M>
.