10.3 Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
10.3.1 Resetting the measurement settings
The following settings for pH measurements are reset to the default settings
with the
Reset
function:
pH
2.
Using
<ENTER>
, activate the
Stability control
function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the measured
parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears and the display of the
measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the
Stability control
function manu-
ally with
<ENTER>
at any time. If the
Stability control
function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3.
Using
<ENTER>
, start a further measurement with
Stability control
.
or
Release the frozen measured value again with
<AR>
.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Buffer TEC
Calibration interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
Resolution pH
0.001
Resolution mV
0.1
Asymmetry
0 mV