Technical data
Multi 3320
96
ba77053e04
07/2018
15.2 General data
Accuracy (± 1 digit)
Variable
Accuracy
Temperature of the test
sample
ϰ
and
ρ
/ temperature compensation
None (Off)
± 0.5 %
Nonlinear (nLF)
± 0.5 %
0 °C ... +35 °C
according to EN 27 888
± 0.5 %
+35 °C ... +50 °C
enhanced nLF function
Linear (lin)
± 0.5 %
+10 °C ... +75 °C
SAL
/ range
0.0 ... 42.0
± 0.1
+5 °C ... +25 °C
± 0.2
+25 °C ... +30 °C
TDS [mg/l]
± 0.5 %
T [°C] / temperature sensor
NTC 30
± 0.1
PT 1000
± 0.1
The accuracy values specified here apply exclusively to the meter.
The accuracy of the measuring cell has also to be taken into
account.
Further data are given in the documentation of your sensor.
Dimensions
ca. 180 x 80 x 55 mm
Weight
Approx. 0.4 kg
Mechanical structure
Type of protection
IP 67
Electrical safety
Protective class
III
Test certificates
CE
Ambient
conditions
Storage
-25 °C ... +65 °C
Operation
-10 °C ... +55 °C
Admissible relative
humidity
Yearly mean: < 75 %
30 days/year: 95 %
Other days: 85 %