Multi 3410
Operation
54
ba75863y18
01/2015
4.6.8
Displaying calibration records
The calibration data can be displayed and then output to the interface.
Displaying the
calibration record
The calibration record of the last calibration is to be found under the
menu item,
Calibration
/
Calibration record
. To open it in the measured
value display, press the
<CAL
_
>
key.
The calibration records of the last 10 calibrations are to be found in the
menu,
Calibration
/
Calibration data storage
/
Display
. To open the
Cal-
ibration
menu, press the
<MENU/ENTER>
key in the measured value
display.
Menu item
Setting/
function
Description
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration
records.
Further options:
Scroll through the calibra-
tion records with
<
><
>
.
Output the displayed cali-
bration record to the inter-
face with
<PRT>
.
Output all calibration
records to the interface
with
<PRT
_
>
.
Quit the display with
<ESC>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to USB
flash drive or printer
-
Outputs the calibration data
storage to the interface USB-
B (USB flash drive/USB
printer)
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the calibration data
storage to the interface USB-
B (PC)
Summary of Contents for wtw Multi 3410
Page 1: ...OPERATING MANUAL ba75863e18 01 2015 Multi 3410 DIGITAL METER FOR DIGITAL IDS SENSORS...
Page 6: ...Multi 3410 6 ba75863y18 01 2015...
Page 14: ...Multi 3410 Overview 14 ba75863y18 01 2015...
Page 18: ...Multi 3410 Safety 18 ba75863y18 01 2015...
Page 106: ...Multi 3410 Operation 106 ba75863y18 01 2015...
Page 116: ...Multi 3410 What to do if 116 ba75863y18 01 2015...
Page 125: ......