Operation
Multi 3410
85
ba75863y18
01/2015
Note
To position the sensor optimally and permanently while it is measuring,
fix it on a support.
4.10.2 Measuring
You can carry out turbidity measurements as follows:
Selecting the
displayed
measured parameter
You can switch between the following displays with
<M>
:
Turbidity [FNU]
Turbidity [NTU]
Freezing the measured
value (HOLD function)
With the HOLD function, you can freeze the current measured value.
The displayed measured value stops changing until you switch the
HOLD function off.
Note
If the HOLD function is active, you can, e.g. start a manual measure-
ment with stability control.
Stability control
(AutoRead )
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values. The display of the measured
1
Perform the preparatory activities according to section 4.10.1.
2
Immerse the sensor in the test sample at an angle and then
position it in the measuring vessel (see section 4.10.1).
30.09.2012 08:00
Info
1
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
2
Release the frozen measured value again with
<AR>
.
The HOLD function is switched off.
The [HOLD] status display disappears.
Summary of Contents for wtw Multi 3410
Page 1: ...OPERATING MANUAL ba75863e18 01 2015 Multi 3410 DIGITAL METER FOR DIGITAL IDS SENSORS...
Page 6: ...Multi 3410 6 ba75863y18 01 2015...
Page 14: ...Multi 3410 Overview 14 ba75863y18 01 2015...
Page 18: ...Multi 3410 Safety 18 ba75863y18 01 2015...
Page 106: ...Multi 3410 Operation 106 ba75863y18 01 2015...
Page 116: ...Multi 3410 What to do if 116 ba75863y18 01 2015...
Page 125: ......